DETERMINATION OF RESIDUAL QUANTITIES OF NEODISHER LABOCLEAN PLM DETERGENT ON LABWARE

Authors

  • N. L. Tarasenko State Institution “O. M. Marzeyev Institute for Public Health of the National Academy of Medical Sciences of Ukraine”
  • N. O. Savina State Institution “O. M. Marzeyev Institute for Public Health of the National Academy of Medical Sciences of Ukraine”
  • V. M. Britsun State Institution “O. M. Marzeyev Institute for Public Health of the National Academy of Medical Sciences of Ukraine”
  • N. V. Оstanina State Institution “O. M. Marzeyev Institute for Public Health of the National Academy of Medical Sciences of Ukraine”

DOI:

https://doi.org/10.11603/2312-0967.2020.1.10903

Keywords:

cleanliness of automated washing of laboratory glassware, detergent, Neodisher Laboclean PLM, disinfection and washing machine, substances that are oxidized, conductivity, specific electrical conductivity, calibration schedule

Abstract

The aim of the work. The development of simple rapid approaches for residual quantities of Neodisher Laboclean PLM detergent in laboratory glassware determining, which are based on the use of pharmacopeia methods (testing the electrical conductivity and the test “oxidizable substances”) and assessing the quality of glassware washing with a disinfection-washing machine.

Materials and Methods. Distilled and deionized water was obtained, respectively, on a GFL2008 aquadistillator and a Millipore Direct-Q 3 UV system. Conductivity studies were performed on a Hanna HI 2300 conductivity meter. Neodisher LaboClean PLM detergent solutions were used. Washing of laboratory glassware was carried out in a G7883 disinfection-washing machine.

Results and Discussion. The electrical conductivity of solutions of various concentrations of Neodisher Laboclean PLM was measured. The obtained data was processed by Excel in order to build a calibration graph and derive its mathematical equation. These approaches, despite the lack of selectivity, make it possible to identify residual contamination of laboratory glassware with Laboclean PLM detergent in concentrations, respectively, of 10-5% and 10-2%. In a similar way, a calibration graph was obtained and the residual amount of one of the hardness salts of tap water (CaCO3) – was estimated. Studies of the cleanliness of laboratory glassware after washing in a washer-disinfector were carried out on the basis of the received data.

Conclusions. The obtained data (0.8–3.2 and 2.3–10.0 μg/dishware, respectively, for Neodisher LaboClean PLM detergent and CaCO3) indicate that the cleanliness of laboratory glassware after washing in a G7883 washer-disinfector is satisfactory for solving of the most tasks of analytical and pharmaceutical chemistry.

Author Biographies

N. L. Tarasenko, State Institution “O. M. Marzeyev Institute for Public Health of the National Academy of Medical Sciences of Ukraine”

lead engineer

N. O. Savina, State Institution “O. M. Marzeyev Institute for Public Health of the National Academy of Medical Sciences of Ukraine”

scientific researcher

V. M. Britsun, State Institution “O. M. Marzeyev Institute for Public Health of the National Academy of Medical Sciences of Ukraine”

DS (Chemistry), chief researcher

N. V. Оstanina, State Institution “O. M. Marzeyev Institute for Public Health of the National Academy of Medical Sciences of Ukraine”

PhD (Economics), scientific researcher, head of the State Scientific Research Laboratory for Quality Control of Medicines

References

United States Pharmacopeia. General Information <1051> USP-NF, Copyright © 2012. The United States Pharmacopeial Convention, 12601 Twinbrook Parkway, Rockville.

Bazel YR, Antal IP, Lavra VM, Kormosh ZhA. Methods for the Determination of Anionic Surfactants. Journal of Analytical Chemistry. 2014;69(3): 211-36. doi: 10.1134/S1061934814010043 DOI: https://doi.org/10.1134/S1061934814010043

Kamaya M. Overview of simple test for determination of surfactants by adhesion method. Journal of Environmental Science and Engineering B. 2013;2: 672-7.

Wang J, Zhiping DU, Wang W, Xue W. Titrimetric determination of anionic surfactant contentin anionic/nonionic surfactant mixture. Turkіsh Journal of Chemistry. 2012;36: 545-55. English. doi:10.3906/kim-1109-42 DOI: https://doi.org/10.3906/kim-1109-42

https://www.drweigert.com/com/uploads/tx_product_manager/downloads/product/safetySheet/neodisher-LaboClean-PLM_SDB_GBen_PN4122_2017-06-08.PDF

The State Pharmacopoeia of Ukraine: in 3 vol. Kharkiv: Ukrainian Scientific Pharmacopoeia Center of Quality of Medicinal Products. Ed.2 [Державна Фармакопея України: в 3 т. / ДП "Український науковий фармакопейний центр якості лікарських засобів" – 2-е вид.], Kharkiv: Ukrainian Scientific Pharmacopoeia Center of Quality of Medicinal Products, 2014; Ukrainian.

The State Pharmacopoeia of Ukraine: in 3 vol. Kharkiv: Ukrainian Scientific Pharmacopoeia Center of Quality of Medicinal Products. Ed.2 [Державна Фармакопея України: в 3 т. / ДП "Український науковий фармакопейний центр якості лікарських засобів" – 2-е вид.], Kharkiv: Ukrainian Scientific Pharmacopoeia Center of Quality of Medicinal Products, 2015; Ukrainian.

Britsun VM, Оstanina NV. [Features of conductometric quality control of distilled water for pharmacopoeial needs]. Pharmacology and Drug Toxicology. 2018;2(58): 97-103. Ukrainian.

Published

2020-02-28

How to Cite

Tarasenko, N. L., Savina, N. O., Britsun, V. M., & Оstanina N. V. (2020). DETERMINATION OF RESIDUAL QUANTITIES OF NEODISHER LABOCLEAN PLM DETERGENT ON LABWARE. Pharmaceutical Review Farmacevtičnij časopis, (1), 42–50. https://doi.org/10.11603/2312-0967.2020.1.10903

Issue

Section

Analysis of drugs